BEAT VOL
Beat volume
Range: 0-10. 0 indicates disabled and 10 indicates the
maximum volume.
•
PACE ON: When ON is selected, a detected pacemaker signal is
indicated by a "|" symbol above the ECG waveform.
OFF: When OFF is selected, the pacemaker analysis is
disabled.
NOTE
•
When monitoring a patient with a pacemaker, PACE must be turned ON.
Otherwise, the system will count the pacemaker pulse as QRS complex. Do not
completely depend on the alarms of heart rate. The patient with a pacemaker must
be nearly monitored.
•
When PACE is turned ON, the system will neither detect the arrhythmia relating
to premature ventricular beats (including PVCs counting) nor perform the ST
analysis. When monitoring a patient without a pacemaker, PACE should be
turned OFF.
•
The PACE tag can be printed in the real-time recording when PACE is turned
ON.
•
NOTCH
Determines whether filter or not.
ON: The monitor protects the signals from the noise
generated by the power line.
OFF: No filtering.
NOTE
•
If the filter method of the ECG waveform is set to a non-diagnostic mode, only the
NOTCH option ON is active and the monitor filters the signals of the power line
frequency; if the filter method is set to the diagnostic mode, two NOTCH options,
both ON and OFF, are active, and the system sets the NOTCH to OFF
automatically.