3.4.6 Auxiliary Output Port
The auxiliary output port can be used to generate analog signals, nurse call signals or
defibrillator synchronization signals.
•
Analog output signals can be generated when the monitor is connected to an
oscilloscope or a pen recorder.
•
If the monitor is connected with the Nurse Call System of a hospital through a special
nurse call cable, the monitor can generate nurse call signals when alarms occur.
•
If the monitor is connected with a defibrillation equipment, the monitor can generate
defibrillator synchronization signals to the defibrillation equipment.
NOTE
•
For detailed connection methods of different uses, please consult the specialist in
your hospital, or our Customer Service.
•
The nurse call cable has two non-polarized conducers at the output end. The
installation should be performed by our servicing engineers or engineers of the
hospital according to the specific nurse call system of the hospital.
WARNING
•
Before defibrillating the patient, the user should ensure the defibrillator and the
monitor have been tested as a system and the two devices can work together safely
and effectively.