A.6 Data Storage
Trend data
Long trend: 96 hours, resolution 1min, 5 min or 10 min.
Short trend: 1 hour, resolution 1 s or 5 s.
Alarm events
70 alarm events and associated waveforms (with user
selectable waveform length 8s, 16 or 32).
ARR events 80 ARR events and associated waveforms with 8s wavelength.
NIBP measurements
800 NIBP groups, including systolic pressures, mean pressures,
diastolic pressures and measurement time.
A.7 Signal Output Specifications
Standards
Meets the requirements of EC60601-1 for short-circuit
protection and leakage current
Output impedance 50.
ECG analog output
Bandwidth (-3dB; reference
frequency: 10Hz)
Diagnostic mode:
Monitor mode:
Surgery mode:
0.05 to 100 Hz (812A module)
0.5 to 40 Hz
1 to 20 Hz
Maximum propagation delay 25 ms (In DIAGNOSTIC mode, NOTCH is OFF)
Sensitivity 1 V/mV± 5%
PACE rejection/enhancement No pace rejection or enhancement
IBP analog output
Bandwidth 0 to 12.5 Hz (-3 dB, reference frequency: 1 Hz)
Maximum propagation Delay 55 ms (the filter function is disabled)
Sensitivity 1 V/100 mmHg ±5%
Nurse call output
Driver Relay
Electrical specifications =60W, =2A, =36VDC, =25VAC
Conducting resistance < 1.
Isolation voltage > 1500 VAC
Signal type Normally open or normally closed, selectable
Defibrillator synchronization pulse
Maximum time delay 35 ms (R-wave peak to leading edge of the pulse)
Amplitude 3.5 V (min) at 3 mA sourcing; 0.8 V (max) at 1 mA sinking
Pulse width 100 ms ±10%
Rising and falling time < 3 ms
VGA
Connector type 15-PIN D-sub socket
Signal RGB: 0.7 Vp-p/75O;
Horizontal/vertical synchronization: TTL level
A-4