APNEA ALM Determines the apnea alarm delay.
If the apnea of the patient exceeds the preset apnea alarm
delay, the monitor triggers an alarm and gives the " CO2
APNEA " alarm message.
Options: 10S, 15S, 20S, 25S, 30S, 35S and 40S.
SWEEP Waveform speed
Options: 6.25, 12.5 and 25.0.
Unit: mm/s.
UNIT Options: mmHg, kPa and %.
WAVE SCALE Options: HIGH and LOW.
This option allows you to adjust the amplitude of the CO2
waveform.
WORK MODE Options: MEASURE and STANDBY.
To start the CO2 monitoring, select the MEASURE mode.
In the STANDBY mode, the air pump and infrared ray
source of the CO2 module are disabled. Hence, decreases
the power consumption and prolongs the operating life of
infrared ray source and the whole CO2 module.
Alarm limits:
Parameter Max. upper limit Min. lower limit Step
CO2 100 0 1
INS 100 / 1
AwRR 150 0 1
Alarm limits in factory configuration:
Parameter Patient type Upper limit Lower limit
Feline 45 30
CO2 Canine 45 30
Other animals 45 30
Feline 4 /
INS Canine 4 /
Other animals 4 /
Feline 100 30
AwRR Canine 100 30
Other animals 100 30
17-6