ALM LEV Alarm level
Options: HIGH and MED.
•
ALM REC Alarm recording
ON: When a SpO2 alarm occurs, the monitor enables the
recording;
OFF: When a SpO2 alarm occurs, the monitor does not
enable the recording.
•
SPO2 ALM HI
SpO2 upper alarm limit
•
SPO2 ALM LO
SpO2 lower alarm limit
•
PR ALM HI
PR upper alarm limit
•
PR ALM LO
PR lower alarm limit
SpO2 and PR alarm limits:
Parameter Max. upper limit Min. lower limit Step
SpO2 100 0 1
PR 254 0 1
The SpO2 and PR alarm limits in factory configuration:
Parameter Patient type Upper limit Lower limit
Feline 100 90
SpO2 Canine 100 90
Other animals 100 90
Feline 180 120
PR Canine 110 60
Other animals 180 120
WARNING
•
Setting the SpO2 upper alarm limit to 100% will disable the upper alarm limit.
High oxygen levels may predispose a premature infant to retrolental fibroplasia.
Therefore, the upper alarm limit for oxygen saturation must be carefully selected
in accordance with the commonly accepted clinical practices.
12-6