Shenzhen PM-9000 Express Manual del operador Pagina 175

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 339
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 174
SWEEP
Waveform speed
Options: 12.5 and 25.0 mm/s.
PR SOUND
Pulse volume
Range: 0–10. 0 indicates the volume is closed and 10
indicates the maximum volume.
SENSITIVE
Sensitivity of SpO2 calculation
Options: HIGH, MED and LOW.
When HIGH is selected, the monitor gives quick response to
the change of the oxygen saturation but the measurement
accuracy might be compromised.
When LOW is selected, the monitor gives slow response to
the change of the oxygen saturation but the accuracy is
relatively high.
NIBP SIMUL ON: When measuring NIBP and SpO2 at the same side, the
SpO2 alarm status remains unchanged until the NIBP
measurement is finished;
OFF: The function of NIBP SIMUL is disabled.
CONFIG >> You can select this option to access the SPO2 CONFIG
menu, in which you may select FACTORY CONFIG or
USER CONFIG. After finishing your selection and exiting
the menu, the system pops up a dialog box asking for your
confirmation.
12-7
Vista de pagina 174
1 2 ... 170 171 172 173 174 175 176 177 178 179 180 ... 338 339

Comentarios a estos manuales

Sin comentarios