To display a 1-millivolt calibration pulse on the ECG wave, select the ECG CAL option in
the ECG SETUP menu. A message “when CAL, can't monitor!” is displayed on the screen.
NOTE
•
If the ECG waveform is too small, or not accurate, and the crocodile clips are
securely and firmly placed to the patient, change the display to a different lead.
11.3 ECG Setup Menu
To display the ECG SETUP menu, use the knob to select the ECG label in the parameter
windows. The following menu appears.
Figure 11-6 ECG SETUP Menu
In this menu, you can perform the following settings:
•
HR ALM Heart rate alarm on/off status
ON: When a heart rate alarm occurs, the monitor gives alarm
indications and stores the alarm;
OFF: When a heart rate alarm occurs, the monitor neither gives
alarm indications nor stores the alarm;
When OFF is selected, the
icon is displayed on the right of
the ECG label.
•
ALM LEV Alarm level
Options: HIGH, MED and LOW.