Shenzhen PM-9000 Express Manual del operador Pagina 316

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 339
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 315
IBP2 NEED
ZERO-CAL
No No 3 The IBP transducer of channel 2
has not been zeroed.
Zero the IBP transducer
of channel 2.
C.2.9 CO Module Alarm Messages
Alarm message A B L Cause Measure
TB SENSOR
OFF
No Yes 3 The cable for blood pressure
measurement is disconnected
from the monitor.
Check for proper
connection of TB cable.
C.2.10 Sidestream CO2 Module Alarm Messages
Alarm message A B L Cause Measure
CO2 NO
WATERTRAP
No No 2 The water trap is not
properly connected or is
disconnected.
Make sure the sidestream
CO2 water trap is firmly
connected.
CO2 SENSOR
TEMP HIGH
No No 1 The temperature of the
sensor assembly is too
high.
Restart the monitor if
necessary. If the problem
remains, contact our
company for repair. CO2 SENSOR
TEMP LOW
No No 1 The temperature of the
sensor assembly is too
low.
CO2 AIRWAY
PRES TOO
HIGH
No No 2 The pressure inside the
airway is too high.
CO2 AIRWAY
PRES TOO
LOW
No No 2 The pressure inside the
airway is too low.
Vista de pagina 315
1 2 ... 311 312 313 314 315 316 317 318 319 320 321 ... 338 339

Comentarios a estos manuales

Sin comentarios